Overcoming High Variation in Your Immunoassay

Consistency is critical to developing a reliable immunoassay. Whether you are optimizing an ELISA, lateral flow assay, or another diagnostic platform, high variation between replicates or across plates can make it difficult to interpret results and determine whether an assay is performing as intended.

High variation can have many causes, from reagent preparation and pipetting technique to incubation conditions and sample matrix effects. The good news is that systematic troubleshooting can help identify the source of variability and improve assay consistency.

What Causes High Variation in an Immunoassay?

When an assay produces inconsistent results, it can be tempting to immediately focus on the reagents or antibodies. However, variation can originate from multiple steps throughout the assay workflow.

Some common contributors include:

  • Pipetting errors: Inconsistent pipetting volumes, poorly calibrated pipettes, or improperly secured tips can introduce well-to-well variability.
  • Reagent preparation: Incomplete mixing, inaccurate dilutions, or contaminated or degraded reagents can affect assay performance.
  • Insufficient reagent homogeneity: Freshly prepared stock solutions and working reagents should be thoroughly mixed to ensure consistent concentrations throughout the assay.
  • Inconsistent incubation conditions: Differences in incubation time, temperature, or agitation can influence binding kinetics and contribute to variation.
  • Plate washing: Inconsistent washing can leave residual material in some wells, contributing to differences in background or signal.
  • Sample matrix effects: Components within patient samples can interfere with antigen-antibody interactions and contribute to inconsistent results.
  • Plate effects: Edge effects, inconsistent environmental conditions, and differences in plate handling can produce spatial variation across a microplate.
  • Bubbles in wells: Air bubbles can interfere with optical measurements and create apparent variation between replicates.

Identifying where variation is introduced is the first step toward improving assay robustness.

Start With Your Standard Curve

If you are seeing high variation in an ELISA, one of the first areas to investigate is the standard curve.

Errors in standard preparation or dilution calculations can affect the entire assay and make it difficult to determine whether observed differences are related to the assay itself or to the standards.

Review your standard curve preparation carefully and confirm:

  • Dilutions were calculated correctly.
  • Standards were thoroughly mixed before use.
  • The appropriate dilution series was prepared.
  • Standards have been stored and handled according to their requirements.
  • Pipetting techniques were consistent across the dilution series.

If there is any question about the accuracy of the standard dilutions, preparing a fresh standard curve can help determine whether preparation error is contributing to the observed variation.

Evaluate Your Pipetting Technique

Small differences in reagent volume can have a significant impact on assay performance, particularly when working with small volumes.

Make sure pipettes are properly calibrated and that pipetting techniques are consistent throughout the assay. Check that pipette tips are securely attached and that the same technique is being used for each replicate.

When using multichannel pipettes, pay particular attention to consistency across all channels. Differences in aspiration or dispensing can create patterns of variation across the plate.

Automating liquid handling, where appropriate, can also help minimize operator-to-operator variability.

Look at Reagent Preparation and Handling

Reagents need to be homogeneous and appropriately prepared before they are added to an assay.

Incomplete mixing can result in different concentrations being delivered to individual wells. Similarly, incorrect buffer or reagent concentrations can affect assay kinetics and signal development.

When troubleshooting high variation, verify:

  • Reagent calculations and dilution factors.
  • Preparation records and expiration dates.
  • Mixing procedures.
  • Reagent storage conditions.
  • Buffer concentrations.
  • Potential contamination.
  • Whether fresh working solutions improve assay performance.

If contamination is suspected, preparing fresh buffers or reagents can help determine whether reagent quality is contributing to the problem.  

Control Incubation Conditions

Temperature, time, and agitation can all influence immunoassay performance.

For example, differences in incubation temperature can affect binding kinetics, while inconsistent incubation times can cause some wells to develop more signal than others. Plate agitation can also be important for ensuring adequate interaction between assay components.

For greater consistency, keep incubation conditions as uniform as possible across the entire plate and from experiment to experiment.

When troubleshooting plate-to-plate variation, pay particular attention to:

  • Incubation temperature.
  • Incubation time.
  • Shaker speed and settings.
  • Plate positioning.
  • Environmental temperature.
  • Plate sealing during incubation.

Consistent environmental conditions can help minimize variation that is unrelated to the biological performance of the assay.

Don't Overlook the Sample Matrix

Even when an assay is carefully optimized, patient samples can introduce another source of variability.

Biological samples contain numerous proteins and other components that can interact with assay reagents. These matrix effects may interfere with target detection, contribute to nonspecific interactions, or affect dilution recovery.

A well-designed sample diluent can help reduce the impact of matrix components while maintaining the intended assay signal.

Surmodics™ IVD offers sample and assay diluents designed to address matrix-related challenges. MatrixGuard™ Diluent and Surmodics™ Assay Diluent (Protein-Free) use different approaches to help reduce matrix interference while maintaining assay performance.

Consider Protein Stability and Blocking

Variation isn't always caused by the steps being performed during the assay. The stability of proteins immobilized on a surface can also affect assay consistency over time.

In addition, insufficient blocking can allow unwanted interactions to occur, increasing background and potentially contributing to inconsistent results.

Surmodics™ IVD's dried protein stabilizers and blockers are designed to stabilize dried proteins while providing blocking functionality in a single step. These reagents can help preserve protein activity and reduce nonspecific binding, supporting more consistent assay performance.

Check for Plate and Edge Effects

If variation follows a pattern across the plate—for example, higher or lower results around the outer wells—edge effects may be contributing to the problem.

Differences in temperature and environmental exposure can affect binding kinetics and signal development. To help minimize these effects:

  • Use a consistent room-temperature environment.
  • Avoid stacking plates during incubation when appropriate.
  • Maintain consistent incubation conditions.
  • Use plate sealers during relevant incubation steps.
  • Use fresh plate sealers and reagent reservoirs as appropriate.
  • Inspect wells for bubbles before reading the plate.

These small process controls can make a meaningful difference when troubleshooting well-to-well or plate-to-plate variation.

A Systematic Approach Can Reduce Variation

When troubleshooting high variation, it can be tempting to change multiple assay components at once. However, changing too many variables makes it difficult to determine what actually improved performance.

Instead, consider taking a systematic approach:

1. Confirm the basics.
Check calculations, pipettes, reagent preparation, expiration dates, incubation times, and temperatures.

2. Examine the pattern of variation.
Is variation occurring between replicates, across the plate, between plates, or between sample types? The pattern can provide clues about the source.

3. Evaluate reagents and samples.
Look at reagent stability, blocking conditions, diluent selection, and potential matrix effects.

4. Change one variable at a time.
A controlled optimization strategy can help identify the factor responsible for the variability.

5. Confirm the improvement.
Once a potential cause has been identified, repeat the experiment to confirm that the change consistently improves assay performance.

Build Consistency Into Your Immunoassay

High variation doesn't necessarily mean an assay needs to be redesigned from the ground up. Often, identifying and controlling a few key sources of variability can significantly improve assay consistency.

From protein stabilization and blocking to sample dilution, Surmodics IVD provides immunoassay reagents designed to help developers address common assay development challenges. The company's portfolio includes protein stabilizers and blockers, sample and assay diluents, detection substrates, antigens/antibodies and other reagents for immunoassay development.

When troubleshooting high variation, start by examining the entire workflow—not just the final signal. A systematic evaluation of reagent preparation, pipetting, incubation, washing, plate handling, and sample matrix effects can help identify the source of variability and move an assay toward more consistent, reliable performance.